Melt Flow Indexer Test Machine

Alarge Melt Flow Index Tester, complies with TS EN ISO 1133 and ASTM D1238 standards, measures the flow and viscosity of polymers under specified conditions. Ideal for determining MFR and MVR values in quality control and production processes.

Standards:

  • TS EN ISO 1133
  • ASTM D 1238

Melt flow index tests are carried out to characterize the melt flow of plastic raw materials, products, and recycled granules in compliance with TS EN ISO 1133 and ASTM D 1238 standards. It simulates the behavior of polymer products or raw materials in the melt phase.

3 models are available: AL MFI-A, AL MFI-A-B, and AL MFI-A-B-O

  • Double-compartment heating system
  • Independent temperature control in both compartments
  • Equivalence in heat distribution
  • User-friendly, high precision with easier use
  • Functional with touch screen
  • Automated cutter – adjustable cutting time
  • Rigid, solid, and stable structure
  • Ground tungsten carbide core (titanium plated) for high precision
  • Automated calculation and reporting software for MFI MVR test
  • Modular structure. AL MFI A, AL MFI A-B, and AL MFI A-B-O instruments are convertible to each other

Standards:

  • TS EN ISO 1133
  • ASTM D 1238
  • Plastic and polymer manufacturing plants
  • Quality control laboratories
  • R&D and material development centers
  • Automotive industry
  • Packaging production
  • Electrical and electronics sector
  • Construction materials manufacturing
  • Medical and healthcare product manufacturing
  • Textile and fabric industry
  • Polymer recycling facilities

Alarge Difference in MFI Test

Precise and Efficient Solution

Modular Structure & Model Transformability

Eriyik Akış Ölçüm Test Cihazı - MFI (MFR & MVR) - 3 Cihaz

Avoid having to buy a new device when your needs change, simply upgrade your first device.

AL-MFI models (AL MFI-A, AL MFI-AB, AL MFI-ABO) feature a modular design and are interchangeable. Transitioning to higher models is easy as needs grow. It reduces investment costs in the long run.

Full Compliance with EN ISO 1133 / ASTM D1238 Standards

“Get your test results globally accepted”

  • AL-MFI device is designed in full compliance with TS EN ISO 1133 and ASTM D1238 standards. This allows laboratories, R&D centers, and quality assurance units to use the device with confidence.

Dual Zone Heating & Adaptive / PID Control System

  • Don’t risk deviation in your test results; forget about temperature imbalances.

  • AL-MFI operates with two independent heating zones (upper/lower) and provides precise temperature control with an adaptive/PID control system. This minimizes temperature deviations in melt flow tests.

High-Precision Orifice and Duct Components

  • Ensure that the sample flow is always controlled.

  • High accuracy components enhance the reliability of flow measurements.

Wide Temperature Range & High Precision

  • Don’t lose temperature control; keep test results consistent.
  • The device can operate in a temperature range of 120 °C to 450 °C. This is a critical feature for accurately measuring the melting behavior of polymers.

Automatic Cutter & Adjustable Cutting Time

  • Do not set the cutting time manually; let the device handle it for you.

  • The AL-MFI device features a step motor-controlled automatic cutting mechanism, and the cutting timing can be adjusted between 0–9999 s.

Sample Weight / Loading Module Options

Eriyik Akış Ölçüm Test Cihazı - MFI (MFR & MVR) - 3 Cihaz
  • Use different sample weights for various polymers to gain flexibility.

    The device supports sample weights ranging from 3 to 8 g. Additionally, with the load module option, you can work with various test weights (from 0.325 kg to 21.6 kg). This allows testing of different raw materials with a single device.

Graphics & Reporting Software, Automatic Calculation

  • Your test results are automatically calculated and reported, making the analysis process easier.

  • AL-MFI Reporting Software offers support for automatic MFI / MVR / MFR calculation and reporting. This accelerates data entry and the analysis process, while minimizing human error.

Powerful Hardware & Durability

  • Your device will operate smoothly for years.

  • Its rigid body structure is made with high-quality components. Models with a net weight ranging between 40–80 kg operate reliably. It can operate with 220V single-phase power.

TECHNICAL SPECIFICATIONS

AL MFI AAL MFI BAL MFI O
Test MethodMFRMFR | MVRMFR | MVR
Orifice MaterialTitanium Coated Tungsten Carbide
Orifice Dimensions2095 ± 0.005 mm
Main Outer Melt ChannelTungsten Carbide
Operating Temperature (Standard Painted Furnace Barrel)120℃ – 300℃
Operating Temperature (Stainless Steel Furnace Barrel)120℃ – 400℃
Temperature Accuracy0.1℃
Temperature ControlAdaptive PID
Heating Time10 min (25℃ – 200℃)
Sample Size3 g – 8 g
Cutting MechanismStep Motor
Cutting Accuracy0.01 Second
Displacement SensorOptionalAvailableAvailable
Displacement AccuracyX5 Microns5 Microns
Automatic Weight LoadingXAvailable
USB Data PortAvailable
Reporting SoftwareAvailable
Net Weight40 kg43 kg80 kg
Power Supply220 V – 50 Hz220 V – 50 Hz220 V – 50 Hz
Power450 W450 W450 W

Frequently Asked Questions

Do you have questions? You're not alone. Here are the frequently asked questions.

The product is valid for 2 years starting from the delivery date.

For your technical support requests, simply contacting us is sufficient. Our expert team will strive to resolve your issue as quickly as possible through remote connection or phone support. If the problem cannot be resolved remotely, our technical team will be dispatched for on-site intervention. This ensures that your device is restored to working condition as swiftly as possible.

The installation and activation of the device you purchased are carried out in your working environment by our expert technical team. After the installation, comprehensive training is provided to your user personnel for the correct and efficient use of the device. This service ensures that you can use all the device’s functions safely and effectively.

To ensure the device operates efficiently and has a long lifespan, regular maintenance is essential. This includes tasks such as cleaning, checking connections, software updates, and calibration. It is recommended to perform maintenance at the intervals specified in the user manual. If desired, you can seek professional support from our technical service for maintenance tasks.

To ensure the device’s accuracy, calibration must be performed at regular intervals. Professional calibration is recommended every 6 months.

Do you have any other questions?

If you still have unanswered questions, please don’t hesitate to contact us. We are happy to provide any additional information you need and are always ready to help!